Dimension Xpand Plus Integrated Chemistry System
Weight | 765 lb (348 kg) Basic 783 lb (356 kg) with Heterogeneous Immunoassay (HM) Module | |
Dimensions | Width: 51” (129 cm) Height: 45” (113 cm) Depth: 31” (79 cm) | |
Operating Temperature | 65-85°F (17-30ºC) with a maximum fluctuation of 5ºF (2.8ºC) per hour. The system requires a maximum of 120 minutes to warm up from a cold start to the incubation temperature. | |
Relative Humidity | 20-80% | |
Avg Noise Output | <70 dBA at 1 m while operating | |
Thermal Output | 3753 BTU/hr (1100 W) | |
Water Usage | Very low water consumption (2 liters/hour at maximum throughput) | |
Power Requirements | ||
Nominal Line | 115 | |
230 | ||
Line Voltage | 103 to 127 | |
207 to 253 | ||
Nominal Line Frequency Hz | 47 to 63 | |
47 to 63 | ||
Maximum Continuous Current, AMPS | ~11 | |
~5.5 | ||
Power Consumption WATTS | 1265 | |
1265 |